Comparison of Cone Beam Computed Tomography and Multi Slice Computed Tomography Image Quality of Human Dried Mandible using 10 Anatomical Landmarks

Samira Saati1, Fatemeh Kaveh2, Shiva Yarmohammadi3

1 Assistant Professor, Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Hamadan University of Medical Sciences, Hamadan, Iran.
2 Dental Student, Department of Community Medicine, Hamadan, Iran.
3 Assistant Professor, Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Kurdistan University of Medical Sciences, Kurdistan, Iran.


NAME, ADDRESS, E-MAIL ID OF THE CORRESPONDING AUTHOR: Dr. Shiva Yarmohammadi, Assistant Professor, Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Kurdistan University of Medical Sciences, Kurdistan, Iran.
E-mail: sh2am3535@yahoo.com
Abstract

Introduction

Cone Beam Computed Tomography (CBCT) has gained a broad acceptance in dentomaxillofacial imaging. Computed Tomography (CT) is another imaging modality for diagnosis and preoperative assessments of the head and neck region.

Aim

Considering the increased radiation exposure and high cost of CT, this study sought to subjectively assess the image quality of CBCT and Multi Slice CT (MSCT).

Materials and Methods

A dry human mandible was scanned by five CBCT systems (New Tom 3G, Scanora, CRANEX 3D, Promax and Galileos) and one MSCT system. Three independent oral and maxillofacial radiologists reviewed the CBCT and MSCT scans for the quality of 10 landmarks namely mental foramen, trabecular bone, Periodontal Ligament (PDL), dentin, incisive canal, mandibular canal, dental pulp, enamel, lamina dura and cortical bone using a five-point scale.

Results

Significant differences were found between MSCT and CBCT and among the five CBCT systems (p<0.05) in visualization of different anatomical structures. A fine structure such as the incisive canal was significantly less visible and more variable among the systems in comparison with other anatomical landmarks such as the mental foramen, mandibular canal, cortical bone, dental pulp, enamel and dentin (p<0.05). The Cranex 3D and Promax systems were superior to MSCT and all other CBCT systems in visualizing anatomical structures.

Conclusion

The CBCT image quality was superior to that of MSCT even though some variability existed among different CBCT systems in visualizing fine structures. Considering the low radiation dose and high resolution, CBCT may be beneficial for dentomaxillofacial imaging.

Keywords

Introduction

Intraoral Two Dimensional (2D) and panoramic radiography are extensively used by dental clinicians for dentomaxillofacial imaging [13]. However, interpretation of 2D images is hard because of the overlapping and superimposition of different structures. Anatomical landmarks such as the lingual foramen and the incisive canal containing neurovascular bundles can be hardly identified or thoroughly assessed on these images. The mandibular bone and the alveolar ridge width cannot be evaluated on these images either [37]. Using, 3D technique has become more popular today [8]. CT has enabled 3D assessment of the craniofacial structures; CT is also available as a diagnostic tool for the head and neck region [9,10] and is ordered prior to some oral surgical procedures [11,12].

It should be noted that CT is not ideal for some diagnostic purposes in dentistry such as evaluation of impacted teeth and periapical lesions. The main disadvantages of CT include high patient radiation dose, limited availability and high cost. CBCT was introduced as an optimal imaging modality due to its sub millimeter resolution, high image quality, short scanning time and decreased patient radiation dose {15 times lower than that of MSCT [13].

Also, CBCT has various applications in oral and maxillofacial surgery for implant placement [1418], bone and tooth fractures [17,19], assessment of temporomandibular joint [20], orthodontic treatment, cases of cleft lip and palate, impacted teeth, osteosynthetic screw [21], third molar extraction [22] and endodontic procedures (assessment of root canal configuration, root resorption and apical lesions) [21,22]. By using appropriate imaging system to pinpoint vital structures, the risk of damage to them during surgery can be reduced [23]. Previous studies comparing MSCT and CBCT showed diverse quality of these systems in showing the details of anatomical structures in the head and neck region [24]. Most previous reports on CBCT and its image quality were limited to one or two systems [2426]. It should be noted that new CBCT systems have been recently introduced to the market.

Thus, considering all the above and also since MSCT has been introduced as a superior imaging technique to visualize the details of bony structures, the aim of this study was to evaluate the image quality of different CBCT systems available in the Iranian market compared to a MSCT system.

Materials and Methods

This experimental study was performed in five private radiology centers and maxillofacial radiology department of Hamadan Dental Faculty, Iran, which lasted for six months.

In this study, dry human mandible was used. The mandible was placed in a plastic container filled with water to simulate the soft tissue. Then, the mandible was scanned by means of five CBCT systems and one MSCT system [Table/Fig-1].

The factor exposure of CBCT systems.

Size of FOV (mm)mAkVp
NewTom 3G6, 9, 120.5110
Scanora 3D7.5×14.5,7.5×10,6×61390
CRANEX 3D6×8,6×41390
Promax8×81284
Galileos15×15490
MSCT25×7.680120

CRANEX® 3D (Soredex, Tuusula, Finland)

New Tom 3G (Quantitative Radiology, Verona, Italy)

Scanora 3D (Soredex, Tuusula, Finland)

Promax (Planmeca, Helsinki, Finland)

Galileos (Sirona, Bensheim, Germany) with multiple Fields of View (FOV)

MSCT (Siemens, Erlangen, Germany)

Image Assessment: Three oral and maxillofacial radiologists evaluated the scans under the same conditions (flat panel monitor; 1440×900 pixels resolution). In order to assess each system, certain software was used: Sygno, Fast Viewer for MSCT, on Demand 3D for Cranex 3D, Romexis Viewer for Promax, Sirona for Galileos and NNT Viewer for New Tom 3G.

They were asked to distinguish 10 anatomical landmarks namely the mental foramen, the mandibular canal, trabecular bone, cortical bone, dental pulp, dentin, enamel, incisive canal, PDL and lamina dura on the scans and then rank them from one to five based on the quality of the images using a five-point Likert scale [22]. The observers were allowed to observe the images in three planes of coronal, axial and sagittal or in each reconstructed plane. Moreover, they were allowed to adjust the contrast and brightness of the images. In order to prevent bias, each radiologist was blinded to the interpretations of others.

Statistical Analysis

Data were analyzed using non-parametric Kruskal–Wallis test. Post-hoc comparisons of different systems were done by the Kendall’s W test. Also, the Friedman’s and Kendall’s W tests were used to assess the intra-observer agreement. Finally, in order to study the inter-observer agreement for the 10 landmarks, two observations were done in total and the results for each landmark were compared in two states. The Cramer’s V was used to compare the assesment by the three radiologists.

Ten landmarks in the mandibular bone were independently observed by three radiologists by means of six different software programs.

In order to rank the quality of images for observation of the landmarks, the five-point Likert scale was used ranging from one (undesirable observation) to five (excellent observation). Each radiologist assessed the landmarks twice. Since, the image quality for observation of the landmarks was reported using a five-point Likert scale, the Kruskal–Wallis and Mann–Whitney U tests (for comparison of average ratings), the Crammer’s V, the Kendall’s tau coefficient and the Friedman’s test were used for calculation of intra and inter observer agreements.

In order to study the differences in image quality among different systems, pairwise comparisons were made using the Mann–Whitney U test.

Finally, to assess the intra observer agreement in different sizes of Field of View (FOV), Wilcoxon signed rank and Friedman’s tests were used and the coefficient of agreement was calculated using the Cochran’s Q and the Kendall’s W tests.

In order to detect the general agreement coefficient of the three radiologists for the 10 landmarks, general reports by each radiologist at two different occasions for the six different systems were analyzed using the Cramer’s V.

Results

For precise evaluation of the scans taken by the different CBCT systems (New Tom, Scanora, CRANEX 3D, Promax, and Galileos) and MSCT in our study, three oral and maxillofacial radiologists evaluated the 10 landmarks (mental foramen, mandibular canal, trabecular bone, cortical bone, dental pulp, dentin, enamel, incisive canal, PDL and lamina dura) twice in a dry human mandible scanned by these systems, and scored the image quality from one to five (excellent=1, good=2, average=3, bad=4 and very bad=5).

Significant differences were noted in the quality of observation of lamina dura (p=0.003), mental foramen (p=0.002), incisive canal (0.000) and PDL (p=0.002); but the difference was not significant for other landmarks [Table/Fig-2].

The factor exposure of CBCT systems.

First RadiologistSecond RadiologistThird RadiologistKruskal–Wallis
Chi- squarep-value
Lamina dura2.763.53.8611.650.003
Mental foramen4.473.594.3612.360.002
Mandibular canal4.5744.455.580.061
Cortical bone4.904.904.595.530.063
Dental pulp4.614.684.591.250.535
Dentin4.524.634.362.290.318
Incisive canal31.684.3630.680.000
Enamel4.614.454.450.7890.674
Periodontal ligament3.283.54.1812.4280.002
Trabecular bone4.574.404.182.2370.327

Among six CBCT systems in two observations (in total), CRANEX acquired the highest score in six landmarks and Promax ranked second in three landmarks. Scanora and Galileos acquired the highest score in one landmark and the MSCT and New Tom acquired the lowest score among all [Table/Fig-3].

The quality score of 10 landmarks given by three radiologist among different systems.

GalileosMSCTPro-maxCRANEX 3DScanoraNew-TomKruskal–Wallis
Chi squarep- value
Lamina Dura3.162.003.504.273.772.2920.1520.001
Mental foramen3.833.165.004.363.94.4 3313.8930.016
Mandible canal4.163.334.665.004.613.9427.357<0.001
Cortical bone5.004.665.005.004.884.5012.4880.029
pulp4.503.005.005.004.944.5535.418<0.001
dentin4.502.004.505.004.944.6133.461<0.001
Incisive canal3.001.833.002.723.272.1113.4350.020
enamel4.661.665.004.814.944.6132.480<0.001
Periodontal ligament3.332.503.834.453.773.5020.0780.001
Trabecular bone4.503.165.004.814.883.7736.917<0.001

Pair wise comparisons of different systems showed that image quality of some systems had significant difference [Table/Fig-4].

Pairwise comparisons to study the differences in image quality among different systems.

Mann-Whitneyp-value
New Tom-Scanora66.50.002
New Tom-CRANEX 3D13<0.001
New Tom-Promax160.011
New Tom-MSCT40.001
New Tom- Galileos430.462
Scanora-CRANEX 3D610.086
Scanora-Promax46.50.614
Scanora-MSCT<0.001<0.001
Scanora-CRANEX 3D610.086
Scanora-Galileos43.50.482
Promax-CRANEX 3D22.50.286
Promax-Galileos13.50.47
Promax-MSCT<0.0010.004

[Table/Fig-5] explained the amount of the intraobserver agreement for different sizes of FOV for each system.

The intraobserver agreement coefficient for different sizes of FOV for each system (2 to 3 FOV sizes).

Friedman and Kendall’sMeanObserver 3
Observer 1Observer 2
Lamina dura1.883.984.2
Mental foramen6.694.255.89
Mandibular canal6.675.36.3
Cortical bone7.6986.57
Dental pulp6.957.56.8
Dentin6.647.365.93
Incisive canal2.191.432.5
Enamel76.826.36
Periodontal ligament2.743.75.36
Trabecular bone6.556.665.09

Chi-square=72.433, Chi-square=123.54, Chi-square=139.31, p<0.001, p<0.001, p<0.001


There was a weak correlation in the radiologists’ interpretation for lamina dura (Cramer’s V=0.293), which was not significant (p=0.62). Similarly, there was a weak correlation for mental foramen (Cramer’s V=0.216) and this correlation was not significant either (p=0.22). The three radiologists had complete agreement in observing the mandibular canal. The three radiologists had complete agreement (100%) in observing the cortical bone as well. There was a strong correlation in the radiologists’ interpretation of dental pulp (Cramer’s V=0.675), which was significant (p=0.01). There was a strong correlation in the radiologists’ interpretation of dentin (Cramer’s V=0.737), which was significant (p=0.003). There was a weak correlation in the radiologists’ interpretation for the incisive canal (Cramer’s V=0.899), which was not significant (p=0.178). There was a strong correlation in the radiologists’ interpretation for the enamel (Cramer’s V=0.737), which was significant (p=0.003). There was a weak correlation in the radiologists’ interpretation for the PDL (Cramer’s V=0.221), which was significant (p=0.204). Also, there was an average correlation in the radiologists’ interpretation for the trabecular bone (Cramer’s V=0.537), which was not significant (p=0.01).

As a whole, it can be said that the difference was significant for four landmarks namely mental foramen, lamial dura, incisive canal and PDL [Table/Fig-6].

General agreement coefficient of the three radiologists regarding the image quality of the 10 landmarks.

Radiologist 1Radiologist 2Radiologist 3Agreement coefficient
Lamina duraGood1318200.293
Bad842
Mental foramenGood1919220.216
Bad230
Mandibular canalGood2121211
Bad000
Cortical boneGood2222221
Bad000
Dental pulpGood2121220.675
Bad010
DentinGood1920200.737
Bad222
Incisive canalGood1720170.889
Bad415
EnamelGood1920200.737
Bad222
Periodontal ligamentGood2018210.221
Bad141
Trabecular boneGood2120210.537
Bad021

Discussion

The use of three-dimensional imaging systems has improved dento-maxillofacial diagnosis [26]. Optimal image quality of vital structures has many benefits for patient and clinician especially in implant therapy.

In the present study, CRANEX 3D had the best image quality in the assessment of 10 landmarks and then stood Promax, Scanora, Galileos and New Tom 3G respectively. MSCT had the lowest image quality. Since New Tom and MSCT have higher kVp than other systems [Table/Fig-1], it can be concluded that high kV and, in turn, higher scattered radiation in these devices result in higher noise of the images which makes the visualization of fine structures difficult. In practice, the choice of CBCT system depends on whatever you need from imaging. In major maxillofacial surgery or orthodontic treatment plan we need Large Field Of View (FOV), so New Tom has the selection criteria but in other treatment planning such as implant therapy the CBCT systems with lower kVp and small FOV seem to be a better option for detecting fine details in vital structures and bone status.

In a study by Liang X et al., it was found that the image quality of CBCT was comparable or even higher than that of MSCT and there were some differences among CBCT systems [27]; in the current study, we concluded that the image quality of CBCT was higher than that of MSCT.

Liang X et al., claimed that bone trabecula and Periodontal Ligament (PDL) were less observable and variable among different systems [27]. In our study, the incisive canal had the poorest quality of observation and PDL did not have a good quality of observation either. This difference may stem from the differences in CBCT systems, which were used in the current study and the study by Liang X et al., as well as the method of image quality evaluation.

Kamburoglu K et al., [28] used landmarks for evaluation of image quality of CBCT and reported different results from the present study. However, regarding the same landmarks (trabecular bone and mandibular canal), both studies concluded that the image quality of these two landmarks in different systems was significantly different.

The study by Hashimoto K et al., was similar to our study and they reported results similar to ours, indicative of higher accuracy of CBCT (3DX) than that of MSCT [22].

Loubele M et al., showed that image quality was better significantly in CBCT than MSCT with respect to lamina dura and PDL observation [29]. They used a dry maxilla for image quality evaluation. Although the systems and the jaw, used in their study were different from the present study but the results were similar.

Lofthag HS et al., evaluated image quality of Accuitomo CBCT in three FOV [30]. Seven radiologists assessed the images. Intra-observer agreement was good and interobserver agreement was moderate. In the present study, interobserver agreement in three landmarks wasn’t good.

Lamina dura was given the least score by the first radiologist and there was a weak agreement among the three radiologists with regard to this landmark as well as mental foramen.

Interestingly, with regard to the mandibular canal and cortical bone, all three radiologists completely agreed (100%) that the image quality of these landmarks was good. The first and third observers gave a low score to the incisive canal, which indicates low image quality of this landmark.

Limitation

However, limitation of this study was the invitro nature of the study and also we didn’t assess image quality of maxilla. Further studies should be carried out for in-vivo evaluation of image quality with other CBCT systems.

Conclusion

The present study concluded that CBCT systems had better image quality and lower noise than MSCT. Among CBCT systems selection of the best device depends on the target landmark that is going to be evaluated.

Chi-square=72.433, Chi-square=123.54, Chi-square=139.31, p<0.001, p<0.001, p<0.001

References

[1]Wenzel A, Digital radiography and caries diagnosis Dento Maxillo Facial Radiol 1998 27(1):3-11.  [Google Scholar]

[2]Tal H, Moses O, A comparison of panoramic radiography with computed tomography in the planning of implant surgery Dento Maxillo Facial Radiology 1991 20(1):40-42.  [Google Scholar]

[3]Farman A G, Farman T T, A comparison of 18 different x-ray detectors currently used in dentistry Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005 99(4):485-89.  [Google Scholar]

[4]Fuchs T, Kachelriess M, Kalender WA, Technical advances in multi-slice spiral CT Euro J Radiol 2000 36(2):69-73.  [Google Scholar]

[5]Bou Serhal C, Jacobs R, Flygare L, Quirynen M, van Steenberghe D, Perioperative validation of localisation of the mental foramen Dento Maxillo Facial Radiology 2002 31(1):39-43.  [Google Scholar]

[6]Sarment DP, Sukovic P, Clinthorne N, Accuracy of implant placement with a stereolithographic surgical guide Internat J Oral Maxillofacial Implants 2003 18(4):571-77.  [Google Scholar]

[7]Dawson P, Lees W R, Multi-slice technology in computed tomography Clinic Radiol 2001 56(4):302-09.  [Google Scholar]

[8]Sheikhi M, Ghorbanizadeh S, Abdinian M, Goroohi H, Badrian H, Accuracy of linear measurments of Galileis cone beam computed tomography in normal and different head position Internat J Dentistry 2012 2012(6):214954  [Google Scholar]

[9]Scarfe WC, Farman AG, Sukovic P, Clinical applications of cone-beam computed tomography in dental practice Journal (Canadian Dental Association) 2006 72(1):75-80.  [Google Scholar]

[10]Hatcher DC, Dial C, Mayorga C, Cone beam CT for pre-surgical assessment of implant sites J California Den Assoc 2003 31(11):825-33.  [Google Scholar]

[11]Madrigal C, Ortega R, Meniz C, López-Quiles J, Study of available bone for interforaminal implant treatment using conebeam computed tomography Medicina Oral Patologia Oraly Cirugia Bucal 2008 13(5):307  [Google Scholar]

[12]Van Assche N, Van Steenberghe D, Guerrero ME, Hirsch E, Schutyser F, Quirynen M, Accuracy of implant placement based on presurgical planning of three dimensional conebeam images: A pilot study J Clin Periodon 2007 34(9):816-21.  [Google Scholar]

[13]Heiland M, Schulze D, Rother U, Schmelzle R, Postoperative imaging of zygomaticomaxillary complex fractures using digital volume tomography Journal of Oral and Maxillofacial Surgery 2004 62(11):1387-91.  [Google Scholar]

[14]Mora MA, Mol A, Tyndall DA, Rivera EM, In vitro assessment of local computed tomography for the detection of longitudinal tooth fractures Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007 103(6):825-29.  [Google Scholar]

[15]Ito K, Gomi Y, Sato S, Arai Y, Shinoda K, Clinical application of a new compact CT system to assess 3-D images for the preoperative treatment planning of implants in the posterior mandible A case report Clin Oral Impl Res 2001 12(5):539-42.  [Google Scholar]

[16]Y Siewerdsen JH, Rafferty MA, Moseley DJ, Jaffray DA, Irish JC, Cone-beam computed tomography on a mobile C-arm: Novel intraoperative imaging technology for guidance of head and neck surgery J Otolaryngol Head Neck Surg 2008 37(1):81-90.  [Google Scholar]

[17]Korbmacher H, Kahl-Nieke B, Schollchen M, Heiland M, Value of two cone-beam computed tomography systems from an orthodontic point of view J Orofac Orthop 2007 68(4):278-89.  [Google Scholar]

[18]Tisklakis K, Syriopoulos K, Stamatakis HC, Radiographic examination of the tempoaramandibular joint using cone beam computed tomography Dento Maxillo Facial Radiol 2004 (33):196-201.  [Google Scholar]

[19]Nakagawa Y, Kobayashi K, Ishii H, Mishima A, Ishii H, Asada K, Preoperative application of limited cone beam computerized tomography as an assessment tool before minor oral surgery Internat J Oral Maxillofacial Surg 2002 31(3):322-26.  [Google Scholar]

[20]Matherne RP, Angelopoulos C, Kulild JC, Tira D, Use of cone-beam computed tomography to identify root canal systems in vitro J Endo 2008 34(1):87-89.  [Google Scholar]

[21]Patel S, Dawood A, Ford TP, Whaites E, The potential applications of cone beam computed tomography in the management of endodontic problems Internat Endo J 2007 40(10):818-30.  [Google Scholar]

[22]Hashimoto K, Kawashima S, Kameoka S, Akiyama Y, Honjoya T, Ejima K, Comparison of image validity between cone beam computed tomography for dental use and multidetector row helical computed tomography Dento Maxillo Facial Radiol 2007 36(8):465-71.  [Google Scholar]

[23]Genu P, Vasconcellos R, Oliveira B, Vasconcelos B, Delgado N, Analysis of anatomical landmarks of the mandibular interforminal region using CBCT in a Brazilian population Braz J Oral Sci 2014 13(4):303-07.  [Google Scholar]

[24]Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, van Steenberghe D, State of the art on cone beam CT imaging for preoperative planning of implant placement Clin Oral Invest 2006 10(1):1-7.  [Google Scholar]

[25]Schulze D, Heiland M, Blake F, Rother U, Schmelzle R, Evaluation of quality of reformatted images from two cone beam computed tomographic systems J Craniomaxillofac Surg 2005 33(1):19-23.  [Google Scholar]

[26]Sheikhi M, Dakhil-Alian M, Bahreinian Accuracy and reliability of linear measurements using tangential projection and cone beam computed tomography Dent Res J 2015 12(3):271-77.  [Google Scholar]

[27]Liang X, Jacobs R, Hassan B, Li L, Pauwels R, Corpas L, A comparative evaluation of cone beam computed tomography (CBCT) and multislice CT (MSCT) Part I. On subjective image quality Euro J Radiol 2010 75(2):265-69.  [Google Scholar]

[28]Kamburoglu K, Murat S, Kolsuz E, Kurt H, Yüksel S, Paksoy C, Comparative assessment of subjective image quality of cross-sectional cone-beam computed tomography scans J Oral Sci 2011 53(4):501-08.  [Google Scholar]

[29]Loubele M, Jacobs R, Maes F, Denis K, White S, Coudyzer W, Image quality vs radiation dose of four cone beam computed tomography scanners Dento Maxillo Facial Radiol 2008 37:309-18.  [Google Scholar]

[30]Lofthag-Hensen S, Thilander-klang A, Grondhal K, A comparative evaluation of cone beam computed tomography (CBCT) and Multi Slice CT (MSCT) Part I. On subjective image quality Elsevier Ireland Ltd 2010   [Google Scholar]