JCDR - Register at Journal of Clinical and Diagnostic Research
Journal of Clinical and Diagnostic Research, ISSN - 0973 - 709X
Biochemistry Section DOI : 10.7860/JCDR/2014/9218.4442
Year : 2014 | Month : Jun | Volume : 8 | Issue : 6 Full Version Page : CC05 - CC07

Validation of Method Performance of pH, PCO2, PO2, Na+, K+ of Cobas b121 ABG Analyser

Sunil Kumar Nanda1, Lopamudra Ray2, Asha Dinakaran3

1 Associate Professor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India.
2 Assistant Professor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India.
3 Tutor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India.


NAME, ADDRESS, E-MAIL ID OF THE CORRESPONDING AUTHOR: Dr. Sunil Kumar Nanda, Associate Professor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Ganapathichettikulam, Kalapet, Pondicherry – 605014, India.
Phone: 09789213729,
E-mail: drsknanda1@gmail.com
Abstract

Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory.

Objectives: Evaluation of method performance of pH, PCO2, PO2, Na+, K+ of cobas b121 ABG analyser.

Materials & Methods: The evaluation of method performance of pH, PO2, PCO2, Na+, K+ of cobas b121 ABG analyser was done by comparing the results of 50 patient samples run on cobas b121 with the results obtained from Rapid lab values (reference analyser). Correlation coefficient was calculated from the results obtained from both the analysers. Precision was calculated by running biorad ABG control samples.

Results: The correlation coefficient values obtained for parameters were close to 1.0 indicating good correlation. The CV obtained for all the parameters were less than 5 indicating good precision.

Conclusion: The new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run on patient samples.

Keywords

Introduction

Validation is associated with ensuring and checking that the product complies with set requirements and satisfies certain defined criteria. The laboratory must demonstrate that their tests/methods are fit for the intended use before application to patient samples [1]. The introduction of a new method or new instrument in addition to the existing instruments is a common occurrence in clinical biochemistry laboratory. The performance of the new method or new instrument should be evaluated thoroughly before its routine use in the laboratory [2,3]. The semi-automated analysers are replaced by completely automated analysers to minimize analytical errors and to improve the turnaround time. The new analyser should have acceptable accuracy, precision, measuring range and should be free from interferents. Blood gas measurements and electrolytes are often performed in Point-Of-Care (POC) settings as well as larger clinical laboratories. Even a slight error in reporting ABG and electrolytes can have serious adverse effects on patient management especially critically ill patients. When a new POC analyser is obtained, the performance of the analyser should be validated. The new analyser was cobas b121 from roche diagnostics and rapid lab ABG analyser from Siemens was the existing ABG analyser in our laboratory which was a validated one. Both the ABG analysers (Cobas b121 and rapid lab – Siemens) were based on the principle of ion selective electrode method. Cobas b121 is a ABG analyser which can measure parameters which are critical for decision making. It calibrates by means of fluid system. The sample volume required is 60 micro liters and the results are available within 50 seconds. The operator intervention is least and most of the analyser operations are automated. The aim of our study was to validate the method performance of pH, PCO2, PO2, Na+, K+ of cobas b121 ABG (Arterial blood gas) analyser in our laboratory with the reference ABG analyser – Rapid Lab (siemens) which was well validated and calibrated.

Materials and Methods

The study period was during April and May 2013 at Pondicherry Institute of Medical Sciences (PIMS). The study was a comparative study.

Inclusion criteria

Only the ABG samples that were sent to laboratory in heparinized syringes and ice packs were included in the study. Only those samples which were received by the laboratory within 10 mins of sample collection were included in the study.

Exclusion criteria

Any ABG sample which did not fulfill the inclusion criteria were excluded from the study. Under aseptic precautions, 2 ml of ABG sample were collected in heparinized syringes from radial artery.

Validation of analyser was done by measuring its accuracy, precision and correlation coefficient. The accuracy was assessed by calculating the bias. Bias is the systematic difference between the expected results obtained by the laboratory’s test method and the results that would be obtained from an accepted reference method. The reference method is the validated method & in our study it was Rapid lab values. 50 patient samples were run on cobas b121 and rapid lab (siemens) ABG analysers simultaneously. The bias was calculated for individual samples run on both the ABG analysers and the mean bias obtained was compared with the Total allowable error as per CLIA (Clinical laboratory improvement amendments) guidelines. Bias was calculated by the following formula:

All control samples were measured on both ABG analysers daily. Every control sample result was plotted on the Levey-Jennings chart and checked for acceptability according to Westgard rules (target value ± 2 standard deviations). If control sample value was within acceptable limits, ABG samples were run and bias was calculated. If results were not within acceptable range, control samples were re-run with fresh control samples. If the control values were still found outside control limits, calibration was performed and control samples were rerun and when found within acceptable limits, only then ABG samples were run and bias was calculated.

Precision was assessed by calculating the Coefficient of variance (CV). CV% is the analytical coefficient of variation of the test method. Coefficient of variance (CV) were calculated from Biorad internal QC (lot number – 22221) for the parameters. Correlation between the analysers were done by regression analysis. Correlation coefficient was calculated from the patient samples run on both the analysers. The values obtained in Cobas b121 and Rapid lab was represented by Bland Altman plot. The parameters validated were pH, PO2, PCO2, Na+ & K+.

Results

The results are represented in Bland Altman plot. The intercept, slope, correlation coefficient for all the parameters are given in [Table/Fig-1]. The correlation coefficient for all the parameters were close to 1 indicating good correlation. The average bias and allowable bias for the parameters are given in [Table/Fig-2]. The observed bias for all the parameters were within the allowable bias as per CLIA guidelines. The Bland Altman plot for pH, PO2, PCO2, Na+& K+ are illustrated in [Table/Fig-3,4,5,6 and 7]. The CV values for the parameters are mentioned in [Table/Fig-8]. The CV values for all the parameters were within 5% indicating good precision.

Intercept, slope & correlation coefficient of parameters studied

ParameterInterceptSlopeCorrelation Coefficient (r)
pH-0.061.000.984
PO2-1.81.030.989
PCO2+1.90.960.984
Na+-10.21.070.969
K+-0.101.020.984

Average bias and allowable bias of parameters studied

ParameterAverage biasAllowable bias
pH0.006 %0.04 %
PO2-7.5%+3SD
PCO20.4 %5 %
Na+0.85 %4.0 %
K+0.16 %0.50 %

Bland Altman plot - pH

Bland Altman plot – PO2

Bland Altman plot – PCO2

Bland Altman plot – Na+

Bland Altman plot – K+

CV% of parameters studied

ParameterCV%
pH0.1
PO22.7
PCO22.4
Na+1.0
K+1.0

Discussion

Documentation of laboratory methods, particularly method selection, method performance and validation has assumed a new significance with the introduction of ISO/IEC 17025. Accreditation bodies, like College of American Pathologists (CAP) and National Accreditation Board for Testing and Calibration Laboratories, India (NABL) require that laboratories validate the performance of tests regularly [4]. Method validation is one of the essential components in achieving Total Quality Management in a new clinical laboratory. Validation ensures that accurate and precise results are reported in a clinically relevant turnaround time [4]. The values of bias and precision obtained from the validation studies can be used to plan the Quality control (QC) protocol for the analyser. According to FDA, validation is “Establishing documented evidence which provides a high degree of assurance that a specific process will consistently produce a product meeting its predetermined specifications and quality attributes” [5]. Each laboratory is responsible for validating the test methods being done in their laboratory. Majority of the manufacturers often support the laboratory in validating the newly installed analyser [4].

In our study, the CVs for all the parameters were less than 5 %. In a validation study done by V.S.Smolcic et al., the parameters that did not meet the required precision were total protein, albumin, calcium, sodium, chloride, Immunoglobulins and HDL cholesterol. In our study, the observed bias for all the parameters were within the acceptable limits suggested by CLIA guidelines. In a validation study done by V.S.Smolcic et al., all the parameters fulfilled the required degree of accuracy except total protein, calcium, sodium and chloride [2]. In our study, the r2 obtained for the parameters pH,PO2, PCO2, Na+, K+ were 0.970, 0.983, 0.972, 0.942, 0.973.The symbol for correlation of determination is by r2. It is the square of coefficient of correlation. The degree of linear correlation of variables is determined by coefficient of determination. In a validation study done by Barnali Das on ALP, r2 observed was 0.99 indicating high degree of linear correlation [4].

Jasna Jurice k et al., did a study on validation of olympus AU2700 plus analyser and observed that precision and accuracy for majority of analytes were within acceptable limits except for sodium and chloride. Instrument was fully comparable with Olympus AU2700 analytical analyser for all analytes expect few analytes like ALT, RF, sodium, CK, total bilirubin and triglycerides showed minor deviations [6]. According to ISO 15189 standard, assigned turn-around time (TAT) is 1 hour for emergency tests. This TAT can certainly be achieved by having two analysers with comparable results. If laboratory results are not comparable from different analytical systems, possibility for incorrect diagnosis and poor quality of patient care is increased [7]. Laboratory errors may be preanalytical, analytical or postanalytical. It has been proved in many studies that preanalytical errors are more common than analytical errors [8]. Analytical errors due to analyser malfunction can be completely avoided by having two comparable functional analysers [7].

Conclusion

Based on the values obtained for correlation coefficient, bias and CV for all the parameters, it can be concluded that the new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run patient samples.

References

[1]Linnet Kristian, Boyd James C, Selection and Analytical Evaluation of Methods-With Statistical Techniques.In:Burtis AC, Ashwood ER, Brurns DE, editors Tietz Textbook of Clinical Chemistry and Molecular Diagnostics 2006 4th editionPhiladelphiaSaunders:353-407.  [Google Scholar]

[2]Vesna Supak Smolcic, Lidija Bilic-Zulle, Elizabeta Fisic, Validation of methods performance for routine biochemistry analytes at Cobas 6000 analyser series module c501 Biochemia Medica 2011 21(2):182-90.  [Google Scholar]

[3]Westgard JO, Darcy T, The truth about quality: medical usefulness and analytical reliability of laboratory tests Clin Chim Acta 2004 346:3-11.  [Google Scholar]

[4]Das Barnali, Validation Protocol: First Step of a Lean-Total Quality Management Principle in a New Laboratory Set-upin a Tertiary Care Hospital in India Ind J Clin Biochem 2011 26(3):235-43.  [Google Scholar]

[5]Food and Drug Administration, Guideline on General Principlesof Process Validation, May 1987  [Google Scholar]

[6]Juricek Jasna, Derek Lovorka, Unic Adriana, Serdar Tihana, Marijancevic Domagoj, Zivkovic Marcela, Analytical evaluation of the clinical chemistry analyser Olympus AU2700 plus Biochemia Medica 2010 20(3):334-40.  [Google Scholar]

[7]Miler M, Simundic A, Stefanovic M, Ferenec-Ruzic D, Kvaternik M, Topic E, A model for results comparison on two different biochemistry analysers in laboratory accredited according to the ISO 15189 Biochemia Medica 2009 19(3):287-93.  [Google Scholar]

[8]Szecsi PB, Odum L, Error tracking in a clinical biochemistry laboratory Clin Chem Lab Med 2009 47:1253-7.  [Google Scholar]