Journal of Clinical and Diagnostic Research, ISSN - 0973 - 709X
Biochemistry Section DOI : 10.7860/JCDR/2021/47818.14722
Year : 2021 | Month : Mar | Volume : 15 | Issue : 03 Full Version Page : BC20 - BC23

Sigma Metrics: A Powerful Tool for Performance Evaluation and Quality Control Planning in a Clinical Biochemistry Laboratory

Trupti Diwan Ramteke1, Anita Shivaji Chalak2, Shalini Nitin Maksane3

1 Assistant Professor, Department of Biochemistry, Seth GS Medical College and KEM Hospital, Parel, Mumbai, Maharashtra, India.
2 Professor and Head, Department of Biochemistry, Seth GS Medical College and KEM Hospital, Parel, Mumbai, Maharashtra, India.
3 Assistant Professor, Department of Biochemistry, Seth GS Medical College and KEM Hospital, Parel, Mumbai, Maharashtra, India.


NAME, ADDRESS, E-MAIL ID OF THE CORRESPONDING AUTHOR: Trupti Diwan Ramteke, 105, Sai Rachana CHS, Plot No. 58/B, Sector 21, Kamothe, Navi Mumbai, Maharashtra, India.
E-mail: drtruptibiochem@gmail.com
Abstract

Introduction

Any error in the laboratory testing processes can affect the diagnosis and patient management. Six Sigma is a data driven quality management system for identifying and reducing errors and variations in clinical laboratory processes.

Aim

This study was carried out to estimate Sigma metrics of various biochemical analytes in order to evaluate performance of quality control and implement optimum quality control strategy for each analyte.

Materials and Methods

This retrospective, observational study was conducted in year 2020 based on the data obtained for a period of six months (July 2019 to December 2019). Sigma metrics for 20 analytes was calculated by using internal quality control and external quality control data. Further, QGI was calculated for analytes having sigma value of <4 to identify imprecision or inaccuracy. Statistical analysis was performed using Microsoft office excel 2010 software.

Results

Total protein, Glucose, Urea, Triglyceride (TAG), High Density Lipoprotein (HDL), and Low Density Lipoprotein (LDL) for normal (L1) and pathological (L2) controls achieved excellent performance (>6 sigma). Westgard rule (13s) with two control measurement (N2) per QC event and run size (R1000) i.e. 1000 patient samples between condivutive QC events was adopted for these analytes. For analytes with sigma value of 4-6, more rules (sigma 4-5: Westgardrules-13s/22s/R4s/41s, N4 and R200 and for sigma value 5-6: 13S/22s/R4s, N2 and R450) were adopted. The sigma values of six analytes (Creatinine, Sodium, Potassium, Calcium, Chloride, Inorganic phosphate) were <4 at one or more QC levels. For these analytes, strict QC procedures (Westgard rules-13s/22s/R4s/41s/6x, N4 and R45) were incorporated. QGI of these analytes was <0.8 which indicated the problem of imprecision. Staff training programs and review of standard operating procedures were done for these analytes to improve method performance.

Conclusion

Sigma Metrics estimation helps in designing optimum QC protocols, minimising unnecessary QC runs and reducing the cost for analytes having high sigma metrics. Focused and effective QC strategy for analytes having low sigma helps in improving the performance of those analytes.

Keywords

Introduction

Laboratory diagnostics plays a crucial role in healthcare setting for accurate diagnosis and patient management. A clinical laboratory should always strive to achieve highest quality goals and provide accurate and reliable test results. Laboratory testing processes can be divided in three phases i.e., preanalytical, analytical and postanalytical. Quality improvement measures can effectively prevent the errors in each phase of the laboratory processes [1]. Quality Control (QC) in the analytical phase is done by using Internal Quality Control (IQC) and External Quality Control (EQC) in order to maintain accuracy and precision of the laboratory results [2]. But, QC cannot detect the number of errors that occur in the laboratory during analytical phase. This justifies the need for adoption of quality management strategies like Six-Sigma for quality assurance. Six-Sigma is a quality management system which was developed by Bill Smith (an engineer in Motorola) in 1980’s to make improvements by identifying errors and mistakes. It is being widely implemented in business, industry and healthcare sector [3].

Six-Sigma is the ultimate benchmark of all processes that can fit in six Standard Deviations (SD) on either side of mean. It is uniform way of describing quality in terms of defects per million opportunities. Six-Sigma performances represent 3.4 defects per million operations [4]. Sigma metrics estimation can effectively evaluate existing QC processes in a laboratory for any shortcomings, and help in selecting optimum QC frequency and multi-rules for each analyte. Thus, it helps in improving accuracy and error detection rate of analytical tests and reducing the false rejection rate [5]. Quality Goal Index (QGI) indicates the problems of inaccuracy and imprecision for the analytes with respect to their quality goals [6]. Various Studies [7-11] have been done to estimate sigma metrics in clinical laboratories to quantitatively evaluate errors and improvise QC strategies. The present study was carried out to estimate Sigma Metrics of various biochemical analytes and to estimate the performance of QC in the Central Clinical Biochemistry Laboratory (CCBL). Further, Quality Goal Index (QGI) was also calculated to identify the problems of inaccuracy and imprecision for analytes having lower sigma values.

Materials and Methods

The present observational study was conducted in the year 2020 at the CCBL of Seth GS Medical College and KEM Hospital in Mumbai after obtaining Institutional Ethics Committee approval (EC/OA-113/2020).

Data was collected retrospectively for a period of six months (July 2019 to December 2019). Sigma Metrics was calculated for biochemical analytes run on two automated AU 680 clinical chemistry analysers (Beckman Coulter, Inc. Brea, California, United States) and on Rx Imola clinical chemistry analyser (Randox laboratories, Crumlin, United Kingdom).

The biochemical analytes assayed on AU 680 were-Bilirubin Direct, Bilirubin Total, Calcium, Chloride, Glucose, Potassium, Total protein, Sodium, Urea, Albumin, Aspartate Transaminase (AST), Alanine Transaminase (ALT), Creatinine, Inorganic phosphate. The biochemical analytes assayed on Rx Imola were: Triglyceride (TAG), Total Cholesterol (TC), High Density Lipoprotein (HDL), Low Density Lipoprotein (LDL), Lactate Dehydrogenase (LDH) and Creatine Phosphokinase (CPK).

Internal Quality Control was performed daily before sample processing on both AU 680 and Rx Imola analysers using QC material (Normal-L1 and Pathological-L2) provided by Beckman Coulter for AU 680 instruments and Randox Laboratories for Rx Imola. The QC practices such as control material storage, reconstitution and analysis were done as per the manufacturer’s instructions. In the laboratory, IQC data was interpreted daily by using Levy Jennings’ charts and Westgard rules. The samples of patients were analysed only when the IQC results were within control limits. The Westgard’s rules 13s, 22s, were considered as rejection rules, and 12s as a warning rule for each analyte [1]. Maintenance and calibration of instruments were done regularly. EQC data was obtained by participating in monthly Randox International Quality Assessment Scheme (RIQAS) and biweekly Lipid cycles.

Statistical Analysis

Statistical analysis was performed using Microsoft office excel 2010 software. Mean, SD for biochemical analytes were calculated using IQC data. Coefficient of variation (CV%) was calculated by using laboratory mean and SD.

CV%=(SD×100)/Laboratory mean

Bias% was calculated by using RIQAS data, for each analyte. Peer group mean is the mean of all QC values of laboratories enrolled in the RIQAS program using the same instrument and method.

Bias%=(Peer group mean-Laboratory mean)/Peer group mean

Total Allowable Error% (TEa%) values of all biochemical analytes were adopted from Clinical Laboratories Improvement Act (CLIA) guidelines [12].

The sigma metrics for analytes was calculated by the following equation:

Sigma=(TEa%-Bias %)/CV%

TEa (observed) was calculated using the following formula:

TEa (Observed)=Bias%+(CV%×2)

Observed TEa was compared with TEa adopted from CLIA.

QGI was calculated by the following formula:

QGI=Bias/(1.5×CV)

QGI was estimated for analytes having less than 4 sigma. If QGI is <0.8, it indicates imprecision. QGI of 0.8 to 1.2 indicates both imprecision and inaccuracy and QGI >1.2 indicates inaccuracy [6].

Results

Sigma values, CV%, Bias%, TEa (observed) of 14 analytes at the QC material L1 and L2 were calculated for AU 680-1 and AU680-2 clinical chemistry analysers and are summarised in [Table/Fig-1,2], respectively. Sigma Metrics of 6 biochemical analytes at the QC material L1 and L2 was also calculated for Rx Imola clinical chemistry analyser and is summarised in [Table/Fig-3].

Sigma values, CV%, bias, and TEa of 14 biochemical analytes obtained using clinical chemistry analyser module AU 680-1.

Sr. No.Name of parameterCV %Bias %TEa (CLIA)SigmaTEa (Observed)
L1L2L1L2L1L2
1Albumin1.21.61.485.54.13.84.6
2ALT2.42.92.3155.34.47.18.1
3AST2.52.72.5155.04.67.57.9
4Bilirubin direct3.63.22.8204.85.4109.2
5Bilirubin total3.33.33.1205.15.39.79.5
6Calcium3.13.51.8102.62.38.08.8
7Chloride1.72.00.652.62.24.04.6
8Creatinine2.52.61.6103.43.26.66.8
9Glucose1.21.40.786.16.63.12.9
10Phosphate inorganic3.63.52.0102.22.39.29.0
11Potassium1.21.40.853.53.03.23.6
12Total protein1.01.20.787.36.12.73.1
13Sodium0.91.10.643.83.12.42.8
14Urea1.31.11.096.27.33.63.2

AST: Aspartate transaminase; ALT: Alanine transaminase; CV: Coefficient of variation; CLIA: Clinical laboratories improvement act; TEa: Total allowable error


Sigma values, CV%, bias, and TEa, of 14 biochemical analytes obtained using clinical chemistry analyser module AU 680-2.

Sr. No.Name of analytesCV %Bias %TEa (CLIA)SigmaTEa (Observed)
L1L2L1L2L1L2
1Albumin1.31.71.285.24.03.84.6
2ALT2.63.12.1155.04.27.38.3
3AST2.42.62.4155.34.87.27.6
4Bilirubin direct3.83.32.4204.65.3109.0
5Bilirubin total3.33.12.9205.25.59.59.1
6Calcium3.13.72.5102.42.08.79.9
7Chloride1.62.10.552.82.13.74.7
8Creatinine2.72.91.3103.23.06.77.1
9Glucose1.21.10.686.26.73.02.8
10Phosphate inorganic3.93.71.8102.12.29.69.1
11Potassium1.11.30.953.73.23.13.5
12Total protein0.91.11.087.86.42.83.2
13Sodium0.91.10.543.93.22.32.7
14Urea1.31.20.896.36.33.43.2

AST: Aspartate transaminase; ALT: Alanine transaminase; CV: Coefficient of variation; CLIA: Clinical laboratories improvement act; TEa: Total allowable error


Sigma values, CV%, bias, and TEa, of 6 biochemical analytes obtained using clinical chemistry analyser module RX Imola.

Sr. No.AnalytesCV%Bias %TEa (CLIA)SigmaTEa (Observed)
L1L2L1L2L1L2
1TAG2.01.81.0157.07.85.04.6
2Total cholesterol1.61.80.9105.75.14.14.5
3CPK3.53.23.9204.65.010.910.3
4LDH3.02.81.5154.54.87.57.1
5HDL2.93.11.2206.56.17.07.4
6LDL3.12.81.1206.16.87.36.7

CV: Coefficient of variation; CLIA: Clinical laboratories improvement Act; TEa: Total allowable error; TAG: Triglyceride; CPK: Creatine phosphokinase; LDH: Lactate dehydrogenase; HDL: High density lipoprotein; LDL: Low density lipoprotein


According to the sigma value, biochemical analytes were divided into 5 categories- excellent, (sigma ≥6), good (sigma 4-6), fair (sigma 3-4), marginal (sigma 2-3), unacceptable (sigma <2) as shown in [Table/Fig-4] [13]. The analytes with sigma value <3 were considered as poor performers.

Sigma metrics performance of various analytes.

PerformanceUnacceptableMarginalFairGoodExcellent
Sigma<22-33-44-6>6
AU 680 1L-1NilCalcium, Chloride, Inorganic phosphorusCreatinine, Sodium, PotassiumAlbumin, ALT, AST, Bilirubin (Total, Direct)Total protein Glucose, Urea
L-2NilCalcium, Chloride, Inorganic phosphorusCreatinine, Sodium, PotassiumAlbumin, ALT, AST, Bilirubin (Total, Direct)Total protein, Glucose, Urea
AU 680-2L-1NilCalcium, Chloride, Inorganic phosphorusCreatinine, Sodium, PotassiumAlbumin, ALT, AST, Bilirubin (Total, Direct)Total protein, Glucose, Urea
L-2NilCalcium, Chloride, Inorganic phosphorusCreatinine, Sodium, PotassiumAlbumin, ALT, AST, Bilirubin (Total, Direct)Total protein, Glucose, Urea
Rx ImolaL-1NilNilNilTotal Cholesterol, CPK, LDHTAG, HDL, LDL
L-2NilNilNilTotal Cholesterol, CPK, LDHTAG, HDL, LDL

AST: Aspartate transaminase; ALT: Alanine transaminase; TAG: Triglyceride; CPK: Creatine phosphokinase; LDH: Lactate dehydrogenase; HDL: High density lipoprotein; LDL: Low density lipoprotein


Total protein, Glucose, Urea on AU 680, TAG, HDL, LDL on Rx Imola for L1 and L2 were the analytes which showed excellent world class performance while the analytes with poor performance (sigma <3) were Calcium, Chloride and Inorganic phosphate.

QGI ratio of biochemical analytes with sigma value <4 (Creatinine, Sodium, Potassium, Chloride, Inorganic Phosphorus, Calcium) for both QC material L1 and L2 was less than 0.8; hence the performance was attributed to problem of imprecision [Table/Fig-5].

Quality goal index analysis for analytes with sigma value <4.

InstrumentAnalyteLevel 1Level 2Problem
AU 680 1Creatinine0.40.4Imprecision
Sodium0.40.4
Potassium0.40.4
Calcium0.40.3
Chloride0.20.2
Inorganic phosphorus0.40.4
AU 680 2Creatinine0.30.3Imprecision
Sodium0.40.3
Potassium0.50.5
Calcium0.50.5
Chloride0.20.2
Inorganic phosphorus0.30.3

Discussion

Sigma Metrics is reliable tool used by clinical laboratories which ensures quality by identifying defects in term of precision and accuracy quantitatively. Sigma performance of 6 is considered as excellent performance. Sigma value of 3 is considered as minimum acceptable level of quality. Sigma performance of an analyte can be used to select appropriate control rules and frequency of runs for a method [8]. In this study, Sigma Metrics of 20 analytes (14 analytes on two AU680 clinical chemistry analysers and 6 analytes on Rx Imola clinical chemistry analyser) was analysed. In the CCBL, Westgard rules followed daily as a part of Statistical QC procedures (SQC) are- 12s/22s/13s with two control measurements (L1 and L2) once daily. These rules are used for all the analytes measured in the laboratory. QC is repeated only when any out of control value is observed for particular analyte after taking corrective actions.

Westgard recently described SQC procedures based on Sigma Metrics-Run Size Matrix and Westgard Sigma Rules with Run Size which includes three parameters: 1) selection of appropriate Westgard Sigma Rules; 2) total number of control measurements per SQC event (N); and 3) frequency of SQC events (Run size (R) of patient samples between SQC events [10,14]. Same was implemented in the biochemistry laboratory to achieve <5% probability of false rejection (Pfr) and ≥90% probability of error detection (Ped) for various biochemical analytes as per new Clinical and Laboratory Standards Institute (CLSI) C24-Ed4, guideline [10].

For analytes with sigma value of >6 (Total protein, Glucose, Urea on both AU 680 analyser, TAG, HDL, LDL on Rx Imola for both L1 and L2),13s rule with two control measurements (N2) per QC event and a run size of 1000 patient samples between QC events (R1000) was adopted [10,14]. However, controls should be analysed every day [14]. Albumin, AST, ALT, Bilirubin (Total, Direct) on both AU 680 for L1, L2 and Total Cholesterol, CPK, LDH for L1 and L2 on Rx Imola were good performers having sigma value of 4-6. Good performers with sigma 5-6 require 13s, 22s, R4s multirules with N2, R 450, and analytes with sigma 4-5 require 13s, 22s, R4s, 41s with N2, R200. For the analytes with sigma value <4 (Creatinine, Sodium, Potassium, Chloride, Calcium, Inorganic phosphorus for one or both QC material levels, multi-rules (13s/22s/R4s/41s/6x) with N4 and R45, were adopted [14]. The SQC procedures for various biochemical analytes according to their sigma value are detailed in [Table/Fig-6].

Statistical Quality Control (QC) procedures selected for biochemical analytes [14].

Sigma of analytesWestgard sigma rulesNumber of control levels (N), run size patient samples between QC events (R)
>613SN2 and R1000
5-613S/22s/R4sN2 and R450
4-513s/22s/R4s/41sN4 and R200
<413s/22s/R4s/41s/6xN4 and R45

As revealed by QGI, the main reason of poor performance of 6 analytes (creatinine, sodium, potassium, calcium, chloride, and inorganic phosphorus) was imprecision. Staff training and review of Standard Operating Procedures (SOP) was done for these analytes to improve method performance. The quality of test results is dependent on various factors such as reagents quality, type and quality of QC materials, types of analysers, methodology followed, environmental conditions, training and personal competency of laboratory staff performing the tests. Hence, during root causes analysis, various aspects associated with methodology, materials, personnel, equipment, working conditions were investigated.

Laboratory staff training (for reagent preparation and control material reconstitution, instrument maintenance, reagent handling, storage), periodical competency assessment program was introduced to improve their attitude and knowledge in order to improve precision for biochemical analytes having low sigma values. Further, SOPs were also reviewed for those analytes having low sigma values and rewritten in simpler and user-friendly manner. In present study, on both AU-680 instruments, sigma performance of 14 biochemical analytes was comparable and analytes which showed sigma value <4 were same on both instruments. It indicates that there was no inter-instrument variability in term of instrument performance. The variation in sigma metrics by various studies can be due to difference in methods of analysis, instruments, IQC material, difference in bias calculated due to different EQC testing bodies [8-11].

Within analyte variation of sigma values of L1 and L2 was seen for some analytes as depicted in [Table/Fig-1,2 and 3]. This can be due to variation in method performance at normal and higher concentration level for particular analyte [10]. Strict SQC procedures should be followed to abolish such variation. The effective QC design based on sigma metrics ensures quality performance by quickly detecting medically significant errors. Fewest number of control measurements for analytes with sigma >6, save cost associated with QC materials, reagents and consumables [13]. Hence, quality improvement and cost reduction can go hand in hand by adopting quality management strategies like Six-Sigma. It serves the quality goal of laboratory to produce accurate results for correct and timely diagnosis which in turn can help in improving health of society by reducing morbidity and mortality.

Limitation(s)

Data to show effectiveness of new SQC procedures designed for all analytes is not included in this paper. A study of sigma metrics analysis with revised and modified QC strategy for analytes having low sigma value is under process.

Conclusion(s)

Six-Sigma is a data driven process improvement system which relies on measuring processes and making improvements by reducing errors and defects. Sigma Metrics helps to identify the shortcomings in the existing QC strategy, determine sources of variation and factors influencing laboratory processes to achieve benchmark quality goals of a laboratory. Six-Sigma metrics is easily adaptable to clinical laboratory and moves the laboratory towards proactivity, quality performance and continuous improvement.

Before this study, common QC strategy was being used for all analytes. Use of optimum QC procedures reduced unnecessary QC runs and associated cost, for analytes with the high sigma metrics result. Focused and effective QC strategy for analytes having low sigma improved the performance of those analytes.

AST: Aspartate transaminase; ALT: Alanine transaminase; CV: Coefficient of variation; CLIA: Clinical laboratories improvement act; TEa: Total allowable errorAST: Aspartate transaminase; ALT: Alanine transaminase; CV: Coefficient of variation; CLIA: Clinical laboratories improvement act; TEa: Total allowable errorCV: Coefficient of variation; CLIA: Clinical laboratories improvement Act; TEa: Total allowable error; TAG: Triglyceride; CPK: Creatine phosphokinase; LDH: Lactate dehydrogenase; HDL: High density lipoprotein; LDL: Low density lipoproteinAST: Aspartate transaminase; ALT: Alanine transaminase; TAG: Triglyceride; CPK: Creatine phosphokinase; LDH: Lactate dehydrogenase; HDL: High density lipoprotein; LDL: Low density lipoprotein

References

[1]Burtis CA, Ashwood ER, Burns DE, Tietz Textbook of Clinical Chemistry and Molecular Diagnostics 2012 5th edSt. Louis, MOElsevier Saunders10.17511/ijmrr.2014.i02.20  [Google Scholar]  [CrossRef]

[2]Shah S, Saini R, Singh SB, Aggarwal O, Goel AK, Six sigma metrics and quality control in clinical laboratory Int J Med Res Rev 2014 28(2):140-49.  [Google Scholar]

[3]Adiga US, Preethika A, Swathi K, Sigma metrics in clinical chemistry laboratory - A guide to quality control Al Ameen J Med Sci 2015 8(4):281-87.  [Google Scholar]

[4]Westgard S, Bayat H, Westgard JO, Analytical sigma metrics: A review of Six sigma implementation tools for medical laboratories Biochem Med (Zagreb) 2018 28(2):02050210.11613/BM.2018.02050230022879  [Google Scholar]  [CrossRef]  [PubMed]

[5]Mao X, Shao J, Zhang B, Wang Y, Evaluating analytical quality in clinical biochemistry laboratory using Six Sigma Biochem Med (Zagreb) 2018 28(2):02090410.11613/BM.2018.02090430022890  [Google Scholar]  [CrossRef]  [PubMed]

[6]Kumar BV, Mohan T, Sigma metrics as a tool for evaluating the performance of internal quality control in a clinical chemistry laboratory J Lab Physicians 2018 10(2):194-99.10.4103/JLP.JLP_102_1729692587  [Google Scholar]  [CrossRef]  [PubMed]

[7]Verma M, Dahiya K, Ghalaut VS, Dhupper V, Assessment of quality control system by sigma metrics and quality goal index ratio: A roadmap towards preparation for NABL World J Methodol 2018 8(3):44-50.10.5662/wjm.v8.i3.4430519539  [Google Scholar]  [CrossRef]  [PubMed]

[8]Maksane S, Parikh R, Vaswani L, Quantitative assessment of analytical phase quality of clinical biochemistry parameters using sigma metrics International Journal of Medical Laboratory 2017 4(2):81-90.  [Google Scholar]

[9]Singh B, Goswami B, Gupta VK, Chawla R, Mallika V, Application of sigma metrics for the assessment of quality assurance in clinical biochemistry laboratory in India: A pilot study Indian J Clin Biochem 2011 26(2):131-35.10.1007/s12291-010-0083-122468038  [Google Scholar]  [CrossRef]  [PubMed]

[10]Zhou B, Wu Y, He H, Li C, Tan L, Cao Y, Practical application of Six Sigma management in analytical biochemistry processes in clinical settings J Clin Lab Anal 2020 34:e2312610.1002/jcla.23126  [Google Scholar]  [CrossRef]

[11]Iqbal S, Mustansar T, Application of sigma metrics analysis for the assessment and modification of quality control program in the clinical chemistry laboratory of a tertiary care hospital Ind J Clin Biochem 2017 32(1):106-09.10.1007/s12291-016-0565-x  [Google Scholar]  [CrossRef]

[12]US. Department of health and human services. Clinical Laboratory Improvement Amendments of 1988 (CLIA) Proficiency Testing Regulations Related to Analytes and Acceptable Performance. 42 CFR Part 493, Federal Register/Vol. 84, No. 23/ Monday, February 4, 2019/Proposed Rules. Pp. 1564-65. Available at- https://www.cdc.gov/clia/clia-documents.html (Accessed on 6th September 2020)  [Google Scholar]

[13]Labs for life project. Training Module. Quality control. Volume:1. Guidelines for common statistical methods used in clinical laboratories. Page no.42. Available at: http://labsforlife.in/reis.aspx?QS=3 (Accessed on 7th September 2020)  [Google Scholar]

[14]Westgard JO, Westgard SA, Establishing evidence-based statistical quality control practices Am J Clin Pathol 2019 151:364-70.10.1093/ajcp/aqy15830517600  [Google Scholar]  [CrossRef]  [PubMed]