Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique

Shadi Khakpour Kermani, Alireza Khatony, Rostam Jalali, Mansur Rezaei, Alireza Abdi

1. Faculty of Nursing, Department of Nursing, School of Nursing and Midwifery, Kermanshah University of Medical Sciences, Kermanshah, Iran. 2. Associate Professor, Department of Nursing, School of Nursing and Midwifery, Kermanshah University of Medical Sciences, Kermanshah, Iran. 3. Assistance Professor, Department of Nursing, School of Nursing and Midwifery, Kermanshah University of Medical Sciences, Kermanshah, Iran. 4. Associate Professor, Department of Health, Kermanshah University of Medical Sciences, Kermanshah, Iran. 5. Assistance Professor, Department of Nursing, School of Nursing and Midwifery, Kermanshah University of Medical Sciences, Kermanshah, Iran.

Correspondence Address :
Dr. Alireza Khatony,
Associate Professor, Nursing and Midwifery School, Dolat Abad Street, Kermanshah- 6714634698, Iran.
E-mail: AKhatony@gmail.com

Abstract

Introduction: The prevalence of diabetes mellitus has magnified, globally. Being cost effective and user friendly, using the glucometer systems has become more widespread. However, despite their advantages, there are controversies regarding their accuracy and precision.

Aim: To determine the accuracy and precision of the common glucometers of Bionime, Accu-check, and Beurer compared with the laboratory technique.

Materials and Methods: In a cross-sectional study, 257 patients of a general internal unit of Imam Reza hospital of Kermanshah were recruited. The blood sample of the subjects was transferred to the laboratory, then, two samples of capillary blood were tested by the three glucometers. The data was entered into a checklist, then entered to SPSS-16.0 software, and analyzed by descriptive and inferential statistics.

Results: In this study, there was a positive correlation between the three blood glucose measurement methods with the standard approach, significantly, which indicates acceptable accuracy. However, this correlation was higher in the Beurer Glucometer (r=0.976, p<0.001). There were no significant differences in repeated measurements of blood glucose in the glucometers of Beurer (t=0.96, p=0.92), and Bionime (t=0.50, p=0.61) by paired t-test, but this test was significant for Accu-chek system (t=2.7, p=0.006).

Conclusion: With respect to the acceptable accuracy and precision of Beurer and Bionime glucometers, it is suggested to use these methods for diabetic patients at homes as well as in the hospital units.

Keywords

Blood glucose self-monitoring, Diabetes complications, Diabetes mellitus

How to cite this article :

Shadi Khakpour Kermani, Alireza Khatony, Rostam Jalali, Mansur Rezaei, Alireza Abdi. ACCURACY AND PRECISION OF MEASURED BLOOD SUGAR VALUES BY THREE GLUCOMETERS COMPARED TO THE STANDARD TECHNIQUE. Journal of Clinical and Diagnostic Research [serial online] 2017 April [cited: 2017 Apr 25 ]; 11:OC05-OC08. Available from
http://www.jcdr.net/back_issues.asp?issn=0973-709x&year=2017&month=April&volume=11&issue=4&page=OC05-OC08&id=9613

DOI and Others

DOI: 10.7860/JCDR/2017/23926.9613


Date of Submission: Sep 03, 2016
Date of Peer Review: Oct 17, 2016
Date of Acceptance: Dec 23, 2016
Date of Publishing: Apr 01, 2017

Financial OR OTHER COMPETING INTERESTS: None.

JCDR is now Monthly and more widely Indexed .
  • PubMed Central® (PMC)New
  • Academic Search Complete Database
  • Chemical Abstracts Service
  • Directory of Open Access Journals (DOAJ)
  • EBSCOhostNew
  • Embase & EMbiology
  • Google Scholar
  • HINARI Access to Research in Health Programme
  • Indian Science Abstracts (ISA)
  • Journal seek Database
  • Open J-Gate
  • Popline (reproductive health literature)
  • SCOPUS
  • www.omnimedicalsearch.com